ROSS 2000
Compact High-Speed Diagnostic. Clarity. Modularity.
A compact, modular, and user-friendly instrument built for high-quality streak imaging across diverse environments. Achieve 36 ps temporal resolution with a large, gated photocathode suited for spectrometer coupling and ultra-fast event capture.
Key Benefits
- Large Temporal Range
Users can choose 12 selectable sweep speeds allowing for a tailored range of factory configured temporal resolutions. - Sub 40 ps Temporal Resolution
Achieve timing clarity with the fastest sweep speed delivering down to 36 ps resolution. - Gated Photocathode Flexibility
18 × 3 mm S20 or S25 photocathode pulsed low-field gated cathode for precise event capture and system safety. - Input Optics
Offner Relay with 1:1 magnification and free of chromatic aberrations using all-reflective optics. Spectrometer coupling compatible with light-lock either through Offner relay or direct to cathode. - Periscope Input
Allows timing fiducial to be projected directly onto cathode. - Sophisticated System Control
Full integration with ROSSApp for calibration, analysis, remote setup, and fit-for-purpose workflow.

Mission-Ready Configuration
Comes standard with a compact streak tube assembly, Micro-Channel Plate, gated 18 × 3 mm photocathode, and custom sweep timing modules. Optional Spectrometer coupling compatible. Designed for lab use and remote operation, with adjustable gain.
| Performance Specifications | |
|---|---|
| Temporal Resolution | 33 ps |
| Photocathode Size | 18 × 3 mm (S20 type), S25 available |
| Spatial Resolution | >50% while imaging 10 Lp/mm reticle |
| Streak Tube Line | Photek STX |
| Gating | Pulsed low-field cathode gating |
| Timing Boards | Custom sweep timing modules per unit for optimized resolution |
| Gain Control | Adjustable via single-stage MCP-based image intensifier |
| Rep Rate Maximum | 1 Hz – Single Shot application |
| Imaging Device | Manta G-504B CMOS |
Resources
Datasheet
Datasheet